In my previous blog post about Light Meters I explained why the ideal detector for measuring light from a LED compatible high NA fiber optic is a detector built into an integrating sphere.
This post describes the best method for taking the measurements from Prizmatix fiber optic experts.
Here is their advice:
The correct way of measuring Power for High NA fibers is to insert the end into the integrating sphere, just past the edge. That way all the light goes in the sphere. Make sure not to insert the fiber too deep as it will block some of the light reflecting off the inside surface.
These are photos of how we measure:
With FC Connectors:
With ferrules:
In addition, the Thorlabs computer program has a “maximum power” box.
When testing the fiber, click the “reset” button (marked red) and then insert the fiber into the sphere (without the adaptor).
The correct power emitted from the Fiber will be in the “max” box (marked blue):
Always make sure the correct wavelength is chosen before measuring.
That is all there is to it.
In my next blog post I will have one more piece of advice on the correct hardware configuration.